Features
Clinical correlation of flagship material pairs comparable to reagent values of internationally top brands
Great batch-to-batch consistency
Flagship Raw Material Pairs
Pair No. |
Catalog No. |
Source |
Recommended Usage |
Platform |
Buffer |
Storage |
Shelf Life |
Purity |
Pair 1 |
BECCA153S104 |
CHO |
Coating |
CLIA |
10mMPB+150mM NaCl+0.1% P300,pH7.4 |
-20℃±5℃ |
3 Years |
≥90% |
BEJCA153S106 |
CHO |
Conjugate |
||||||
Pair 2 |
BECCA153S105 |
CHO |
Coating |
|||||
BEJCA153S106 |
CHO |
|||||||
Other |
BECCA153S106 |
CHO |
Coating |
Pairing Antigen
Catalog No. |
Source |
Application |
Buffer |
Storage |
Shelf Life |
GENCA153N101 |
Native |
Immunoassay/Calibration |
0.9% NaCl solution+ protective protein |
-20 °C ± 5 °C |
3 Years |
Performance
Clinical Correlation
Fapon CA153 Flagship Raw Material Pair 1
1. The clinical correlation between Fapon CA153 Pair 1 and the reagent of international top brand R within the concentration range of 0–1600 U/mL is 0.96.
2. The clinical correlation between Fapon CA153 Pair 1 and the reagent of international top brand R within the concentration range of 0-50 U/mL is 0.93.
Fapon CA153
Flagship Raw Material Pair 2
1. The clinical correlation between Fapon CA153 Pair 2 and the reagent of international top brand R within the concentration range of 0–1600 U/mL is 0.98.
2. The clinical correlation between Fapon CA153 Pair 2 and the reagent of international top brand R within the concentration range of 0-50 U/mL is 0.92.
Hook Effect
No false-negative were
found due to hook effect in the concentration of 3000 U/mL of Fapon
CA153 flagship pair 1 & 2 to reagent values of international top brand, which means the false negative rate can be reduced
effectively.
Batch-to-batch Consistency
The
performance of batch-to-batch consistency was validated using the same process
for 3 batches of Fapon CA153 antibodies.
The results showed a activity variation level within 15% between each batch, confirming
high batch-to-batch consistency and reliable product qualities.